Installation Conditions

Synchron Table Light und Synchron Table 4.0

Room conditions for CRUSE scanners

  • The floor in the room must be stable and free of any vibrations.
  • The room walls, ceilings and windows must be painted dark (black / dark gray) or overcast. No foreign light source should be visible during scanning.
  • The room temperature must be between 20 and 25 ° C. Since the room can heat up considerably with frequent scanner use (especially with LED lamps), the room must be air-conditioned.
  • The humidity should be a maximum of 70%.
  • The environment should be as dust-free as possible.
  • Power supply: self-protected current path 230 / V / 16A, 3 x 2.5 mm² with earthing.
  • Direct connection with 3 x 2.5mm² or CEE.
  • Sockets grounded.

Configuration Scanner PC

  • State of the art workstation
  • 2 x Gigabit network cards (one for the scanner and one for the home network)
  • Monitor 24 inch or larger (to display accurate colors the monitor needs to be calibrateable, e.g. Eizo ColorGraphic Series)
  • Windows 10 / 64 bit
  • Connection to the internet
  • Adobe Photoshop (recommended)

Configuration Factory/Hybrid PC

  • High-performance state of the art workstation
  • Minimum 64 GB RAM (recommended 128 GB for SynchronTable 295)
  • High-performance graphic card with CUDA
  • Monitor 24 inch or larger (to display accurate colors the monitor needs to be calibrateable, e.g. Eizo ColorGraphic Series)
  • Windows 10 / 64 bit
  • Connection to the internet
  • Adobe Photoshop (recommended)